INFANT RADIANT WARMER

SPECIAL FEATURES:

Advance computerized Neonatal Resuscitation Device with a large Graphic LCD dual actual temp. display for Baby and Air, Set temp, audio, visual alarms and warnings .
High quality Ceramic heater element with stainless steel parabolic reflector and the protection guard.
Rendered with LED bar for heater ON %,while maintaining set value.
Fitted with special halogen/LED examination light for operating the baby for treatment.
Baby bed can easily be tilted towards both sides for operating procedure, low density washable mattress fit to baby bed in canopy.
MRW-02 having computerized system have facility to operate the same manually or time based treatment.

TECHNICAL SPECIFICATION :-

WIDTH

90cm

DEPTH

110cm
HEIGHT185cm
MATTRESS SIZE 70cm x 50cm x 4cm
MONITOR TRAY SIZE 27cm x 32cm
HEATERCeramic 650W
HEATER ROTATION 90 deg on both side
REFLECTOR Parabolic Reflector
OBSERVATION LAMP 50W Hassle free halogen lamp
SIDE PANELS Unbreakable Polycarbonate drop down
CASTORSStrong Castor wheel with 2 brake
MATTRESS Waterproof, Washable
MATTRESS DENSITY 25Kg/m3
IV POLEAdjustable Height S.S. rod
SENSOR Pt100

 

TEMPERATURE / CONTROLS :-

SUPPLY VOLTAGE190V-250 VAc
SUPPLY FREQUENCY50-60Hz
POWER CONSUMPTION750W
TEMPERATURE DISPLAYCelsius / Fahrenheit
TEMPERATURE RANGE25C – 38 C
TEMPERATURE RESOLUTION0.1 C
TEMPERATURE ACCURACY0.2C
SET TEMPERATURE RANGE20-39 C in 0.1C increment
HEATER OUTPUT0-100% in 1% increment
MODES OF OPERATIONServo(Baby,Air), Manual
TIMER0 to 99 min
MANUAL MODE TIMER0 to 99 min
ALARMSAudio and Visual

 

ALARMS:-

TEMPERATURE HIGHIf Temp > Set Temp by 1 deg C
TEMPERATURE LOWIf Temp < Set Temp by 1 deg C
OVERHEATIf Temp > 38 deg(Baby) / 39 deg (Air)
SENSOR ERRORSensor disconnect/ not working
POWER FAILUREIf power fails
SYSTEM FAILUREIf system fails
HEATER FAILUREIf heater stops