INFANT RADIANT WARMER
SPECIAL FEATURES:
- Advance computerized Neonatal Resuscitation Device with a large Graphic LCD dual actual temp. display for Baby and Air, Set temp, audio, visual alarms and warnings .
- High quality Ceramic heater element with stainless steel parabolic reflector and the protection guard.
- Rendered with LED bar for heater ON %,while maintaining set value.
- Fitted with special halogen/LED examination light for operating the baby for treatment.
- Baby bed can easily be tilted towards both sides for operating procedure, low density washable mattress fit to baby bed in canopy.
- MRW-03 having under surface LED phototherapy which is integrated with open care unit for multi therapy, computerized system have facility to operate the same manually or time based treatment.
TECHNICAL SPECIFICATION :-
WIDTH |
90cm |
DEPTH |
110cm |
HEIGHT | 185cm |
MATTRESS SIZEÂ | 70cm x 50cm x 4cm |
MONITOR TRAY SIZEÂ | 27cm x 32cm |
HEATER | Ceramic 650W |
HEATER ROTATIONÂ | 90 deg on both side |
REFLECTORÂ | Parabolic Reflector |
OBSERVATION LAMPÂ | 50W Hassle free halogen lamp |
SIDE PANELSÂ | Unbreakable Polycarbonate drop down |
CASTORS | Strong Castor wheel with 2 brake |
MATTRESSÂ | Waterproof, Washable |
MATTRESS DENSITYÂ | 25Kg/m3 |
IV POLE | Adjustable Height S.S. rod |
SENSORÂ | Pt100 |
TEMPERATURE / CONTROLS :-
SUPPLY VOLTAGE | 190V-250 VAc |
SUPPLY FREQUENCY | 50-60Hz |
POWER CONSUMPTION | 750W |
TEMPERATURE DISPLAY | Celsius / Fahrenheit |
TEMPERATURE RANGE | 25C – 38 C |
TEMPERATURE RESOLUTION | 0.1 C |
TEMPERATURE ACCURACY | 0.2C |
SET TEMPERATURE RANGE | 20-39 C in 0.1C increment |
HEATER OUTPUT | 0-100% in 1% increment |
MODES OF OPERATION | Servo(Baby,Air), Manual |
TIMER | 0 to 99 min |
MANUAL MODE TIMER | 0 to 99 min |
ALARMS | Audio and Visual |
ALARMS:-
TEMPERATURE HIGH | If Temp > Set Temp by 1 deg C |
TEMPERATURE LOW | If Temp < Set Temp by 1 deg C |
OVERHEAT | If Temp > 38 deg(Baby) / 39 deg (Air) |
SENSOR ERROR | Sensor disconnect/ not working |
POWER FAILURE | If power fails |
SYSTEM FAILURE | If system fails |
HEATER FAILURE | If heater stops |